Section 3 Installation
3-24
For refrigerated baths:
Ensure all communication and electrical connections are made
prior to starting the circulator.
• Install the supplied RJ45 shielded cable between the immersion
circulator and the bath RJ45 connectors (similar to Ethernet). This is
required for proper operation.
RJ45 Connector
On refrigerated baths,
this cable is required
for proper operation
between the bath and the
immersion circulator.
RJ45 Connector
• Install the power cord from the connector on the rear of the controller,
A, to the connector on the rear of the refrigerated bath, B.
• Connect the bath's power cord, C, to a grounded power outlet.
For refrigerated baths, never connect controller power inlet, A, to
a power outlet. Never connect power outlet, B, to anything but an
immersion circulator.
Ensure the electrical cords do not come in contact with any of the
plumbing connections or tubing.
C = Power Inlet from
power supply
Immersion Circulator
A = Power Inlet from Bath
Immersion Circulator
Bath
B = Power Outlet to
Immersion Circulator
Bath