Setup
Start Up
Thermo Scientific
Leave refrigerated baths in an upright position at room temper-
ature (~25°C) for 24 hours before starting. This will ensure the
lubrication oil has drained back into the compressor.
Before starting, double check all USB (optional), electrical and
plumbing connections.
Ensure the circulator's HTC is set to 25°C below the fire point
of the selected fluid.
D o not run the circulator until fluid is added to the reservoir. Have
extra fluid on hand. If the circulator does not start refer to Section 6
Troubleshooting.
• On refrigerated baths, place the circuit protector
located on the rear to the I position.
• Place the circuit protector located on the back of the immer-
sion circulator to the I position.
After a slight delay the blue LED on the front panel illuminates.
• Press
goes out.
• Ensure the start symbol has a highlight box around it, if not
use the arrows to navigate to the symbol.
20.0°C
Highlighted
Start Symbol
• Press
into a stop symbol (
20.0°C
Stop Symbol
Note The pump starts immediately but refrigerated baths take up to
30 seconds before the compressor starts.
Note After start up, check all the plumbing connections for leaks.
, the Start Display appears and the blue LED
SP1
Menu
24.2°C
Start Display
. The circulator starts and the start symbol turns
).
Section 4 Operation
Reservoir
In
Fluid
Temperature
I
O