Bath Circulator
Thermo Scientific
Circulators are designed for continuous operation; and for indoor use.
The equipment normally ships with the immersion circulator mounted facing
the reservoir. You may change the position ±90° by removing the thumb
screws, no tools are required.
Do not mount the circulator backwards; the line cord could contact the
reservoir fluid.
Rotate immersion
circulator only ±90°
from shown position
Never place the circulator in a location where excessive heat, moisture,
inadequate ventilation, or corrosive materials are present.
Leave refrigerated baths in an upright position at room temperature
(~25°C) for 24 hours before starting. This will ensure the lubrication oil
has drained back into the compressor.
A45HC units utilize a flammable refrigerant, risk of fire or explosion,
do not use sharp objects to clean evaporator coil in the tank.