Installation; Ambient Conditions; Ventilation - Thermo Scientific ADVANCED Serie Manual Del Usuario

Ocultar thumbs Ver también para ADVANCED Serie:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Ambient
Conditions

Ventilation

Thermo Scientific

Installation

Section 3
Ambient Temperature Range
Maximum Relative Humidity
Operating Altitude
Overvoltage Category
Pollution Degree
Degree of Protection
*decreasing linearly to 50% relative humidity at 40°C (104°F)
Never place the circulator in a location where excessive heat, moisture,
inadequate ventilation, or corrosive materials are present.
Carefully install the circulator to ensure it does not fall into the bath or
that its line cord does not make contact with the bath contents.
All circulators except A45HC can operate with 0 clearance on two exhaust
sides as long as the third exhaust side has unrestricted air flow. A45HC must
have at least 25mm [1"] of clearance on both sides to comply with UL leak
mitigation safety standards. Blocked ventilation increases the circulator's
temperature, reduce its cooling capacity and, on refrigerated baths, eventually
lead to premature compressor failure.
5°C to 40°C (41°F to 104°F)
80% for temperatures up to 31°C (88°F)*
Sea Level to 2000 meters (6560 feet)
II
2
IP 20
Ventilation Options
3-20
Tabla de contenido
loading

Tabla de contenido