Section 4 Operation
Note Ensure the RTA is set to 0 before doing a calibration.
1. To calibrate the temperature sensor highlight
Calibration
and press
Internal RTD
External RTD
Menu
3. Press
to display:
Internal RTD
High
Low
Calibrate
Menu
5. Depending on which end you are calibrating
High
Low
highlight
or
RTD
temperature stabilizes for several minutes
enter the temperature displayed on your reference
thermometer and press
Internal RTD
High
Low
Calibrate
Menu
7. Repeat for the low end setpoint.
Internal RTD
High
Low
Calibrate
Menu
4-54
to display:
xx.x
xx.x
xx.x
SP
xx.x
Internal
. Once the
.
xx.x
xx.x
xx.x
SP
xx.x
xx.x
xx.x
xx.x
SP
xx.x
2. With the desired sensor highlighted, press
to display:
Calibrate
Restore User Cal
Save User Cal
Restore Factory Cal
Menu
SP
4. Highlight the
temperature box and enter
either the desired high end or low end setpoint
value and press
.
Internal RTD
High
Low
Calibrate
Menu
Calibrate
6. Highlight
complete the high end procedure.
Internal RTD
High
Low
Calibrate
Menu
Once complete you can store the calibration into
Save User Cal
memory by selecting
and pressing
.
You can later restore the same calibration by high-
Restore User Cal
lighting
Another option is to restore the factory calibra-
tion values by highlighting
Cal
and pressing
.
xx.x
xx.x
xx.x
SP
xx.x
and then press
to
xx.x
xx.x
xx.x
SP
xx.x
and pressing
Restore Factory
Thermo Scientific
.