Care Of The Push Button Chuck - Dentsply Sirona Midwest Stylus Plus Instrucciones De Uso

Ocultar thumbs Ver también para Midwest Stylus Plus:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

6607290.book Seite 45 Montag, 5. September 2016 11:53 11
®
Operating Instructions Midwest Stylus
Plus Air-Driven-High-Speed Handpiece
7 Conditioning
1. Fit the spray adapter onto the nozzle of the spray can.
2. Insert the handpiece as far as it will go and hold it in place.
3. Spray into the handpiece for 1-2 seconds.
IMPORTANT: Hold the spray can upright.
4. Wipe any spray that comes out with a disinfection cloth.
5. Repeat the process until the spray is clear.
7.4.2

Care of the push button chuck

®
Sirona
New
Use Midwest
Plus Aerosol Spray on the push button chuck to remove
deposits and ensure proper functioning of the clamping system.
Interval
Interval
● At least once a week
45