8.3

Servicing Sirona quick coupling

8.3.1

Replacing O-rings

1. Remove the defective O-ring.
Using T1 Spray
O-Ring Removal/Attachment _ Using Tool
2. Slide the tool (A), with the new O-ring attached, to just in front of the
corresponding groove (B).
3. Insert the O-ring. Do not use any sharp objects.
4. Lightly oil the O-ring with Midwest
NOTICE! Do not use Vaseline or silicone grease on the O-rings.
®