6607290.book Seite 40 Montag, 5. September 2016 11:53 11
7 Conditioning
40
Operating Instructions Midwest Stylus
1. Check whether the handpiece is clean after conditioning under good
lighting (min. 500 lux) and color rendering index (min. 80 Ra).
2. If it is still dirty, repeat the process.
For further processing, the handpiece should be dry and free of
residue.
3. Blow the handpiece out with max. 3 bar.
4. Maintain mechanical parts manually [ → 43].
5. Maintain the push button chuck manually [ → 45].
6. Pack the handpiece in packaging material suitable for sterilization
and storage. e.g. paper/plastic composite packaging.
7. Sterilize the adapter [ → 47].
®
Plus Air-Driven-High-Speed Handpiece
6607290 D3673.201.01.01.02