6607290.book Seite 58 Montag, 5. September 2016 11:53 11
8 Maintenance
A
B
C
D
58
Operating Instructions Midwest Stylus
7. Note the position of the bulging side (C) when fitting the sealing
washer.
8.3.5
Replacing the felt strip (quick coupling F)
Replace the felt strip if the oil-laden return air that escapes is soiled. We
Replacing the felt strip
recommend replacing the felt strip at least every three months.
1. Click the control ring into the maximum water flow.
2. Place the coupling upright on a firm, flat surface supported by the
screw-on cap and firmly push the control ring (C) downward until it
disengages.
NOTICE! Note that the balls (A) and springs (B) may drop out.
3. Replace the contaminated felt strip (D).
4. Click the control ring back into position, taking care to keep the balls
and springs in place.
®