36
Message
Possible Cause(s)
INVALID
An invalid kit lot barcode was
KIT LOT
detected by the BD Veritor™
BARCODE
Plus Analyzer.
OPERATOR ID
SCAN TIMEOUT
No barcode was scanned
REINSERT
within 30 seconds.
TEST DEVICE
SPECIMEN ID
SCAN TIMEOUT
No barcode was scanned
REINSERT
within 30 seconds.
TEST DEVICE
LOT NUMBER
SCAN TIMEOUT
No barcode was scanned
REINSERT
within 30 seconds.
TEST DEVICE
The shelf or operational life
of the BD Veritor™ Plus
Analyzer is nearing
expiration. The maximum
XXX DAYS
shelf-life of the BD Veritor™
ANALYZER LIFE
Plus Analyzer from the date
REMAINING
of manufacture is 34 months
(regardless of time of first
use). The maximum number
of months from first use is
24 months.
The BD Veritor™ Plus
Analyzer is nearing its usage
XXX READS
expiration. The maximum
REMAINING
number of tests performed by
a BD Veritor™ Plus Analyzer
is 10,000 runs.
Failure of the internal battery.
Battery Not Charging:
In the event the battery icon
does not indicate the battery
is charging while the power
Flashing Plug icon
supply is plugged into the
BD Veritor™ Plus Analyzer,
examine the AC Adapter
plugged into the Analyzer.
The supplied AC Adapter
has a BD label near the
coaxial power connector and
conforms to the sample
labeling in section 2.3.
Corrective Action(s)
Ensure that the barcode scanned during this step is from
a valid BD Veritor™ Test Kit box label. For kits with two
barcodes, scan only the upper barcode that starts with
(17).
Reinsert test device and scan all barcodes within
30 seconds of the display prompt.
Reinsert test device and scan all barcodes within
30 seconds of the display prompt.
Reinsert test device and scan all barcodes within
30 seconds of the display prompt.
Contact your distributor to order a new BD Veritor™ Plus
Analyzer*.
The duration of the message is two seconds. To clear the
message immediately, insert a test device to begin a new
run.
Contact your distributor to order a new BD Veritor™ Plus
Analyzer*.
The duration of the message is two seconds. To clear the
message immediately, insert a test device to begin a new
run.
Contact BD*.
Note: the Analyzer may continue to be used while
plugged into a facility power source.