BD Veritor™ Plus Analyzer Instructions for Use
When the system encounters an alert or error condition, the result number and message is usually displayed.
Contact numbers for BD technical services and local BD representatives outside of the United States appear on page 215.
Result Number
SYSTEM ALERTS
01
POSITIVE CONTROL LINE
02
03
05
06
07
08
09
11
12
* outside of the United States contact your local BD representative
32
Message
Varies according to test
Successful test execution
type
The positive assay control
INVALID
line is outside the normal
CONSULT MANUAL
range.
TEST WAS
The test device was removed
REMOVED
while the run was in
REINSERT
progress.
TEST DEVICE
BACKGROUND INVALID
The assay background area
CONSULT MANUAL
is outside of the normal
ERROR 05
range.
ANALYZER
FAILURE
A hardware anomaly has
CALL BD
occurred.
ERROR 06
ANALYZER
FAILURE
A hardware anomaly has
CALL BD
occurred.
ERROR 07
ANALYZER
A high ambient light
FAILURE
condition has been detected
CALL BD
or a hardware anomaly has
ERROR 08
occurred.
ANALYZER
FAILURE
A hardware anomaly has
CALL BD
occurred.
ERROR 09
ANALYZER
FAILURE
A hardware anomaly has
CALL BD
occurred.
ERROR 11
NEGATIVE CONTROL
Cross-reacting material
LINE INVALID
present.
CONSULT MANUAL
Possible Cause(s)
None required.
On first occurrence, repeat the assay. On subsequent
occurrences, contact BD*.
That run is invalidated. The device may be re-inserted to
begin a new run.
Remove the device and cycle the instrument power.
•
If the instrument displays an error during power-on,
contact BD*.
•
If no error message appears, obtain an additional
sample and repeat the assay. If the issue recurs,
contact BD*.
Contact BD*.
Contact BD*.
On first occurrence, move the unit to a lower ambient light
condition.
In case of persistent recurrence, contact BD*.
Contact BD*.
Contact BD*.
On first occurrence, repeat the assay. On subsequent
occurrences, contact BD*.
Corrective Action(s)