BD Veritor™ Plus Analyzer Instructions for Use
3.1.4

Test Device Slot

The test device slot is located on the right side of the BD Veritor™ Plus Analyzer (see arrow in figure below). Test devices are inserted by
aligning the insertion arrow on the device with the insertion arrow on the Analyzer. A distinctive "click" will be noted when the test device is
properly engaged in the Analyzer.
3.1.5

USB Port

The BD Veritor™ Plus Analyzer includes a USB On-The-Go port which is used for four applications:
Analyzer Update: the BD Veritor™ Plus Analyzer can be updated, using a USB flash drive that is provided by BD, to add capability to
read additional BD Veritor™ assays and to add new features. Instructions to perform these updates are provided with the USB flash
drive when such updates are available.
Printing: An optional printer is connected to the BD Veritor™ Plus Analyzer via this port. See section 2.4 for instructions.
Data Download: BD Veritor™ Plus Analyzers that are equipped with a barcode scanner can be unlocked to enable secure transfer of
the test record to a computer via this port. See section 4.8 for instructions.
Data Transmission: BD Veritor™ Plus Analyzers that are equipped with the BD Veritor™ Plus Connect accessory can transmit test
results to a facility LIS over a secure connection via this port.
The BD Veritor™ Plus Analyzer can be externally powered only from the coaxial power port on the rear of the instrument.
4

Operation

4.1

General

Section 4.2 provides instructions for routine operation of the BD Veritor™ Plus Analyzer using the Analyze Now operational mode. In this
mode, the test device is prepared and developed outside of the Analyzer, then inserted to perform the final read. Section 4.3 provides
instructions for routine operation of the BD Veritor™ Plus Analyzer using the Walk Away operational mode. In this mode, the test device is
prepared outside the Analyzer but allowed to develop inside the Analyzer. The Analyzer initiates the final read at the appropriate time. Both
sections include instructions for operation with or without a BD Veritor™ barcode reader equipped module installed. Sections 4.4 and 4.5
describe the Quality Control (QC) test procedure in Analyze Now and Walk Away mode, respectively.
4.2
Using the BD Veritor
The BD Veritor Plus Analyzer is intended for use with BD Veritor™ Test Devices. Specific instructions for performing a BD Veritor™ Test
can be found in the package insert included with each test kit. After following those instructions the procedure for interpreting a test result
with the BD Veritor™ Plus Analyzer using Analyze Now operation is as follows:
1
Place the BD Veritor™ Plus Analyzer on a flat, dry, stable surface.
2
Ensure that the BD Veritor™ Plus Analyzer is not in direct sunlight or exposed to a bright light.
3
Check the BD Veritor™ Plus Analyzer for an inserted device. If a device is present, remove it from the Analyzer.
20
The Analyzer cannot be powered through its USB port.
Plus Analyzer – Analyze Now Operation
NOTE