BD Veritor™ Plus Analyzer Instructions for Use
Other
Overvoltage Category
Lifetime Specification
Performance Specifications
Compatible 1D Barcode Protocol
Operator ID Maximum Barcode Length
Specimen ID Maximum Barcode Length
Data Transmission Security Protocols
Default Workflow Configuration
Operator Identification Scan
Specimen Identification Scan
Kit Lot Information Scan
Display Language
USB Port
2.6
BD Veritor™ InfoScan Module Setup – Optional
Addition of the BD Veritor™ InfoScan module enables the Analyzer to capture barcoded information and integrate it into the test record.
Utilized together with BD Veritor™ Plus Connect, and the BD Synapsys™ Informatics Solutions it allows secure transmission of test results
to a facility LIS. The included USB cable may be used to network the Analyzer either to a printer or to the BD Veritor™ Plus Connect.
Steps for installing the module are:
1
Turn the Analyzer power off.
2
Turn the Analyzer upside down.
Ensure that the power button is not accidentally pressed while the Analyzer is upside down.
3
Depress the locking tab on the module bay cover.
ELECTRICAL EXPOSURE HAZARD. DO NOT TOUCH THE ELECTRICAL CONTACTS INSIDE THE MODULE BAY.
16
Power to the Analyzer should remain off during module installation.
WARNING
II (per IEC 60664)
™
The BD Veritor
barcode reader equipped modules do
not expire.
™
The BD Veritor
Plus Connect does not expire.
Code 128, Code 39, UPC / EAN, Interleaved 2 of 5, Codabar,
GS1 DataBar, GS1-128. ASCII-128 character set.
20 characters (Longer barcodes are truncated to 20 characters)
20 characters (Longer barcodes are truncated to 20 characters)
HTTPS with SSL/TLS certification. AES-256 payload encryption.
Disabled
Enabled
Disabled
English
Locked
CAUTION