Column Storage
Column Storage
Columns which may be stored in water or buffered
solvents:
CN columns
Ion exchangers
Aqueous SEC packings
However:
However:
Prevent microbial growth by using 0.05% sodium azide
o
in mobile phase
Small quantity of organic solvent (acetonitrile 5% or
o
methanol 10%)