Página 6
Mezcla de 9 componentes que permiten verificar de forma exhaustiva sistemas LCMS o MS en un amplio rango de métodos y condiciones. Mezcla desarrollada por Waters en el centro de I&D y producción de Manchester, y utilizada para la verificación de la instrumentación del Laboratorio de demo.
Página 13
Neutral compounds for most analytical 186006360 HPLC/UPLC assays Reversed-Phase Suitability Neutral, Acid, Base for most 186006363 Standard analytical HPLC/UPLC assays Preparative Chromatography Neutral, Acid, Base for most UV and MS 186006703 purification HPLC assays AutoPurification Dye Mix Visible markers for fraction collection...
Página 14
– moderate peak width Blue analyte: Slowest – BROADEST/MOST DILUTED BAND -- Waters c 1998 Broadest Peak (longer on the column – takes more mobile phase to sweep it out of the column -- broader the peak)
Página 15
Where Band Spreading Occurs Where Band Spreading Occurs Simple Isocratic HPLC System Simple Isocratic HPLC System Mobile Phase HPLC Column (Stationary Phase) Data Data Injector AutoSampler Sample Pump Detector Where Band spreading occurs Waste {Flow path of sample from injector to the detector}...
Página 77
How does flow stoppage cause this problem How does flow stoppage cause this problem with an HPLC column? with an HPLC column? •Analytes •Analytes •Stopped flow with no pressure on the •At flow with pressure on the mobile phase, pores de-wet. Restart flow, mobile phase.
Página 83
•Changes in Retention/Resolution Method Development •Please remember: •HPLC Columns contain packing materials which can be which can be chemically altered chemically altered depending depending on how they are used and treated (mobile phase, pH, ion-pairing reagents, surfactants, age). •“Column History”...
Página 99
Contaminated mobile phase Contaminated mobile phase •4 different buffer as A solvent •(b) •(a) •HPLC grade water as A solvent •Blank gradients 5-80% ACN after •Blank gradients 5-80% ACN after 10min (a) and 30min (b) equilibration 30min (a) and 30min (b) equilibration w.
Página 109
Troubleshooting Concerns Regarding Vials Regarding Vials Vials Can Present both Chemical and Mechanical Problems Most important: What HPLC system is using the vial. Sometimes the autosampler appears to be at fault, but in reality it’s the vial. — Differences in the designs of HPLC systems determine the...
Página 110
•Sample Vials Vials are the least expensive component of an HPLC system, but can be the biggest contributor to problems that you encounter, such as: •Chemical Mechanical Sample draw volume reproducibility Evaporative loss Septum dislodging/Coring Spurious peaks/extractables Injector damage/Robotics Sample/analyte degradation...
Página 129
Column Lifetime Column Lifetime HPLC columns have a finite lifetime. Most common cause of column failure is due to high pressure, typically caused by particulates. Sources of particulate contamination — Debris from instrument seals — Particles in the sample —...
— Usar sales de alta calidad para preparar los tampones — Usar solventes orgánicos de grado HPLC gradiente o superior — — Limpiar con agua después de haber trabajado con tampones Limpiar con agua después de haber trabajado con tampones —...