• Direct contact of prophy powder with surfaces and marginal
areas of dental restorations should be avoided.
ACCESSORIES
• Cavitron Prophy-JET Sodium Bicarbonate Prophy Powder,
130003
• Cavitron JET-Fresh Aluminum Trihydroxide Prophy
Powder, 13020301
• O-Ring Replacement Kit, 62605
INFECTION CONTROL PROCEDURES
WARNINGS
• Instruments are not sterile upon receipt and must be sterilized
prior to use in accordance with the following instructions.
• The use of cleaning and disinfectant solutions is not
recommended.
LIMITATIONS ON PROCESSING
• Repeated cleaning and sterilization cycles have minimum
effect on these instruments. End of life is normally determined
by wear and damage due to use.
• Do not use chemical disinfectants prior to sterilization or rapid
deterioration of the material may occur.
• Cold liquid disinfection/sterilization, chemical vapor
sterilization, and dry heat sterilization methods have not been
tested or validated for efficacy and are not recommended for
use.
INSTRUCTIONS
Point of use:
Remove excess soil with disposable cloth or paper.
Containment and Transportation:
Protect instruments from damage. Instruments should be
reprocessed as soon as is reasonably practical following use.