Function Inspection
To check if the oscilloscope function system of the instrument is
working normally, please perform the following function inspection
steps:
1. Press the Power button to turn on the instrument and it switches
to scope function automatically after the power-on self test.
2. Connect the probe adapter plug to the USB port of the
instrument.
3. Connect the BNC end of the probe to the BNC connector of the
instrument.
4. Connect the alligator clip and probe of another end to the grounding
end and compensation signal output terminal of the adapter plug.
5. Press AUTO to display square waveforms with 1KHz frequency
and about 3V peak value in normal conditions as shown below:
6. Test channel 2 in the same method. If waveforms displayed do
not appear like waveforms above,
Compensation".
RSHS800
14
please perform the "Probe