Cleaning Inspection
Disinfection
Symmetry Sur
terminally sterilized prior to surgical use. See
Sterilization instructions.
Maintenance
Between uses, lubricate moving parts with
a water-soluble lubricant in accordance with
the manufacturer's instructions. PRESERVE
Lubricant is recommended for instruments
to be steam sterilized (REF 43-1033).
Inspection and Functional Testing
210-8011-85-
210-8011-85-H-INS.indd
Load instruments so that hinges are open
and cannulations and holes can drain.
Place heavier instruments on
the bottom of containers. Do not
place heavy instruments on top
of delicate instruments.
For instruments with concave surfaces,
such as curettes, place instrument with
the concave surface facing downward
to facilitate draining.
Inspect all instruments before sterilization
or storage to ensure the complete
removal of soil from surfaces, tubes and
holes, moveable parts.
If areas are diffi cult to inspect visually,
check for blood by immersing or fl ushing
the instrument in a 3% hydrogen
peroxide solution. If bubbling is observed,
blood is present. Rinse instruments
thoroughly after using hydrogen
peroxide solution.
If soil is still present, reclean
the instrument.
gical instruments must be
Visually inspect the instrument and
check for damage and wear.
Cutting edges should be free of nicks
and have a continuous edge.
Jaws and teeth should align properly.
Moveable parts should have smooth
movement without excessive play.
Locking mechanisms should fasten
securely and close easily.
Long, thin instruments should be free
of bending and distortion.
6