4. Press <Right Arrow> key. Press <Up/Down Arrow> keys to scroll to select desired Pipe
5. Press <ENTER> key to save Pipe Size selection.
Select Liquid Class
1. Press <Left Arrow> key to select [Pipe Data] and then press <Down Arrow> key to select
2. Press <Right Arrow> key to select [Liquid Class].
3. Press <Right Arrow> key again to select [Select Liquid]. Press <Right Arrow> key again.
4. Press <Up/Down Arrow> keys to scroll to desired liquid.
5. Press <ENTER> key to save selection.
FST020 IP65 NEMA 4X Quick Start
Operating Instructions, 03/2012, A5E03086489-05
①
Colon must appear first before Pipe Class can be selected.
Size.
Note
Pre-programmed Pipe Size and relevant pipe parameters will appear in menu cells. Enter
dimensions manually if pre-programmed dimensions do not match application.
Note
The DN sizes listed in the [Select Pipe Size] menu option list are referenced to DIN Table
2448. After selecting pipe size, check pipe OD and wall thickness for correct dimensions.
[Application Data].
①
Colon appears.
②
Select new liquid from option list.
Connecting
3.2 Programming the Transmitter
19