Guide and manufacturer's statement - electromagnetic immunity
The TE 780 PLUS electro stimulator is designed to work in the electromagnetic environment described below.
The customer or user of the TE 780 PLUS electro stimulator must guarantee that it used in such an environment.
Immunity test
Electrostatic discharge (ESD)
IEC 6100-4-2
Transiting / fast electric
trains
IEC 61000-4-4
Overvoltage
IEC 61000-4-5
Voltage drops, brief cuts
and variations in voltage
on supply input lines
IEC 61000-4-11
Magnetic field at mains
frequency (50/60 Hz)
IEC 61000-4-8
NOTE: U
is the AC mains voltage rate prior to application of the test level.
T
Recommended separation distances between portable
and mobile radio frequency equipment and the TE 780 PLUS electro stimulator
The TE 780 PLUS electro stimulator is designed to work in an electromagnetic environment where irradiated RF interfe-
rence levels are kept under control. The customer or the user of the TE 780 PLUS electro stimulator may help avoiding elec-
tromagnetic interference by guaranteeing a min distance between mobile and portable RF communication devices (tran -
smitters) and the TE 780 PLUS electro stimulator, as recommended below and to suit the max output power of the radio
communication devices.
Max output
power of the
specified transmitter
(W)
0,01
0,1
1
10
100
Transmitters with a max output power not indicated above: the recommended separation distance in metres (m) can be
calculated using the applicable equation for the transmitter frequency, where P is the max nominal output power of the
transmitter in Watts (W) as indicated by the manufacturer of the transmitter and d is the recommended separation
distance in metres (m). NOTE 1: At 80 MHz and 800 MHz, the highest frequency interval is applied.
NOTE 2: These guidelines may not apply to all situations. Electromagnetic propagation is influenced by the absorption
and reflection of structures, objects and people.
IEC 60601
test level
± 6kV contact
± 8kV in air
± 2kV for power lines
±1kV for input/output
lines
±1kV differential mode
±2kV common mode
5% U
(>95% drop in
T
U
) for 0.5 cycles
T
40% U
(60% drop in
T
U
) for 5 cycles
T
70% U
(30% drop in
T
U
) for 25 cycles
T
<5% U
(>95% drop in
T
U
) for 5 seconds
T
3 A/m
Separation distance at the transmitter frequency (m)
150kHz - 80MHz
1,17
0,12
0,37
1,17
3,7
11,7
92
Conformity
level
Floors must be wooden, concrete
± 6kV contact
or tiled. If floors are covered with a
synthetic material, the relative
± 8kV in air
humidity should be at least 30%.
The TE 780 PLUS electro stimulator
Not applicable
is a device with DC power supply
from an incorporated battery.
The TE 780 PLUS electro stimulator
Not applicable
is a device with DC power supply
from an incorporated battery
The TE 780 PLUS electro stimulator
Not applicable
is a device with DC power supply
from an incorporated battery
The TE 780 PLUS electro stimulator
Not applicable
is a device with DC power supply
from an incorporated battery
80MHz - 800MHz
1,17
0,12
0,37
1,17
3,7
11,7
Electromagnetic
environment - guide
800MHz - 2,5GHz
1,17
0,23
0,74
2,33
7,37
23,3