Incidental surface contact of the optional wireless foot control, power supply and cord with water, soap or a water-based hospital-level disinfection solution
will not damage the material of construction. Do not allow any solution to penetrate the casing.
The instructions provided above have been validated by Dentsply Sirona as being capable of preparing a medical device for re-use. It remains the
responsibility of the processor to ensure that the processing is actually performed using equipment, materials and trained personnel in the processing facility
to achieve the desired result. This requires validation and routine monitoring of the process. Likewise, any deviation by the processor from the instructions
provided should be properly evaluated for effectiveness and potential adverse consequences. Use of other reprocessing procedures/methods is at the
discretion and sole responsibility of the practitioner.

5. TECHNICAL SPECIFICATIONS

Power Supply manufacturer:
Power Supply model number:
AC Input
AC Input Current
AC Input Phases
AC Input Frequency
AC Voltage Range for Power supply
DC Output Power
DC Output Voltage and Current
Output Regulation
Weight
Dimensions
Remote Communication
Transmit/Receive
Operating Environment
Transport and Storage Conditions
Optional Wireless Foot Control
Applied Parts

6. CLASSIFICATIONS

Type of protection against electric shock:
Degree of protection against electric shock:
Mode of operation for Handpiece:
Mode of operation for Foot Control:
Settings for Handpiece
Degree of safety of application in the presence of a
flammable anesthetic mixture with air or with oxygen or
nitrous oxide:
According to medical device directive:
Pollution Degree Classification
Overvoltage Category
14
GLOBTEK WR9QA1200USBNMEDRVB
Continuous (100-240 VAC)
Less than 0.7A
Single
50-60 Hz
120 – 240 V
5W
+5VDC at 1A
+/- 10%
Handpiece with cartridge cap = 120 g
Foot Control = 200 g
Handpiece with cartridge cap: L = 190 mm, W = 30 mm
Optional Wireless Foot Control: W = 118 mm H = 40 mm
Frequency:
Power:
Channels:
Bandwidth
Ambient temperature:
Relative Humidity:
Atmospheric Pressure:
Altitude:
Ambient temperature:
Relative Humidity:
Atmospheric Pressure:
Altitude
Protection Class IPX1.
Single-use disposable polyethylene barrier sleeve
Single-use disposable cartridge mixtip/intraoral tip assembly
Class II
Type B Applied Part
Operating, Charging, OFF
OFF, Continuous
3 (Fast Set, Regular Set, User Set)
Equipment not suitable for use in the presence of flammable mixtures
I (Rule 12) (IEC 60601) UL 60601-1
Pollution Degree 2
Category II (connected to wall outlet)
2405-2480 MHz
1mW
40 (BLE)
2 MHz
10°C - 40°C
5-95% (non-condensing)
80-106 kPa
< 2,000 meters
-20°C to 50°C
5-95% (non-condensing)
50-106 kPa
< 5,000 meters