8. Dispenser has a built-in safety mechanism to prevent cartridge over pressurization. If material is partially or completely set and/or cartridge is
blocked, plungers will slow or retract. Manually retract plungers by simultaneously pressing and releasing Slow ► and Fast ►► speed buttons, and
replace cartridge/mix tip assembly.

3.9. Dispenser disassembly/plunger retraction

1. Prior to disassembly, ensure Dispenser plungers are fully retracted.
• Simultaneously press and release both Slow ► and Fast ►► speed forward speed buttons on the handpiece to retract plungers.
• Plungers will automatically retract as outlined above without additional user input.
4. HYGIENE & MAINTENANCE
Cross-contamination - Infection
• Do not reuse single use products. Dispose of in accordance with local regulations.
• The barrier is designed for single use and must be disposed of after each use in accordance with local regulations.
The barrier is not a replacement for cleaning, disinfection, and sterilization.
• Reprocess reusable products as described below.
NOTICE
Wrong cleaning or disinfection method.
Damage to the device.
1. Always disassemble digit Power
2. Do not clean or disinfect with aggressive agents (e.g. solutions based on orange oil or acetone).
3. Start reprocessing within 1 hour after use.
4. Repeated liquid contact may damage labels. Vigorous wiping can destroy the label. Wipe device labels gently.
For Manual Cleaning and Disinfection
• Discard used gloves according to local regulations.
• Disinfect hands with an appropriate bactericidal, virucidal, and fungicidal hand disinfectant solution according to local regulations. Use according to
disinfectant solution manufacturer's Instructions for Use.
• Use a new clean pair of examination gloves.
INSTRUCTIONS FOR CLEANING, DISINFECTING AND STERILIZING CARTRIDGE CAP
Warnings
• These instructions are for use ONLY on the Cartridge Cap. The handpiece and charging base should be disinfected
• Steam autoclaving sterilization is appropriate for the Cartridge Cap.
• Do not allow the device to exceed 137° C.
Limitations on
• Repeated processing has minimal effect on the cartridge cap.
• The cartridge cap can be subjected to the manual cleaning and disinfection process up to 800 times.
Reprocessing
• The cartridge cap can be subjected to the automated washer/disinfector process up to 800 times.
• The cartridge cap can be subjected to the steam autoclave sterilization process up to 800 times.
• Cold liquid immersion disinfection/sterilization, chemical vapor sterilization, and dry heat sterilization methods have not
Point of Use
• Remove protective barrier and discard according to local regulations.
• Remove excess soil with disposable cloth / paper wipe in combination with an alcohol-based, tuberculocidal, quaternary
• Reprocess dispenser handpiece as outlined in the "Instructions for Cleaning and Disinfecting Other Parts" section.
Prior to Cleaning and Disinfection of the Cartridge Cap, the digit Power
It is recommended that the Cartridge cap and handpiece be reprocessed as soon as is reasonably practical following use.
2. To disassemble: Press both side cartridge cap locks to release.
Withdraw cartridge cap/cartridge/mix tip straight off.
Do not twist the cartridge cap while removing.
Slightly depress the retaining tab and withdraw the used cartridge/mix tip assembly.
3. To prepare/reprocess contaminated components for reuse, see Hygiene & Maintenance Section 4.
Dispenser before final reprocessing.
®
according to the procedures in the "Instructions for Cleaning and Disinfecting Other Parts" section.
been tested or validated for efficacy and are not recommended for use.
ammonium solution with a label claim for cleaning (e.g., VoloWipes
1.
Remove cartridge cap: Press both side cartridge cap locks to release. Withdraw cartridge cap/cartridge/mix tip
straight off. Do not twist the cartridge cap while removing.
2. Slightly depress the retaining tab and withdraw the used cartridge/mix tip assembly. Do not attempt to refill or
consolidate used cartridges. Dispose of used cartridge/mix tip assembly in accordance with local regulations.
Disinfecting/Cleaning/Deodorizing Wipes).
®