58
„AVS-Center" - „Method/Results"
4.7
The tab "Method/Results" is the central input place for measuring parameters, administration of
measuring methods, the selection of the viscometer as well as for the evaluation and protocolling/
logging of the results.
Fig. 9 AVS-Center, "Method/Results"
Here, the measuring parameters are listed on the left and the sample parameters and the evaluation
results on the right.
Under measuring parameters, the number by the "Device" indicates the number of the respective
measuring point. All settings in this tab "Method/Results" refer to this measuring point. In order to
switch measuring points, you need to activate the desired measuring point on the "Overview" tab and
switch back to the tab "Method/Results".
4.7.1 Measurement parameters
4.7.1.1 Operator
Here, the user will log in using his user ID and his password, if you are working with user
administration (see section 4.10).
Depending on the user level, you can enter data here and use the keys. If you have user level 2, there
are no restrictions. For level 1 users, you may merely select measuring levels, enter parameter values,
and select viscometers and measure. You cannot change the measuring parameters under
"Measurement methods" nor can you change the evaluation (formulas).