Safety device
This is the capacitive sensor VZ 8552, which is highly recommended during suctioning operating
modes to monitor the safety bottles VZ 7021 or VZ 7022. The safety bottle is installed in the pneumatic
hose connection between the ViscoPump II and the capillary tube to detect an inadvertent
oversuctioning of the sample into the ViscoPump. This capacitive sensor VZ 8552 will emit an error
signal if the ViscoPump suctions liquid all the way into the safety bottle due to a malfunction.
The sensor is connected to the front panel of the ViscoPump II (refer also section 2.3.4).
Wastebottle
This is the overflow safety system (scale) VZ 8551 for the waste bottle, which is used if a waste
system is connected. (refer also section 2.3.5).
„AVS-Center" - „Overview"
4.6
Select the tab "AVS-Center"-> "Overview", and the read devices will be displayed on the screen.
®
Fort the AVS
370 they will represent the existing modules ViscoPump II.
Fig. 8 AVS-Center, „Overview"
There will be a table displayed for each device, which will look as follows:
Equipment: Address (00-08), status message, start/stop key and measuring values 1 to max. 10.
By selecting "show activity status", the operating status of the viscometer will be displayed in the
bottom part of the window. The direction of the black triangle in the stylized measuring sphere shows
whether the sample is being pumped up or drained. The little red squares are displayed if the liquid is
above or below the top light sensor.