Accessing the Instrument Pages

A single, quick button press turns ON the backlight for 10 seconds.
The Information page can be accessed by pushing the TEST button for
about one second.
The instrument beeps twice.

For All Versions (Excluding the Steel Mill Version):

1) Test gas mode
2) Minimum Oxygen concentration ("LO") - for oxygen only
3) Peak Reading
• Toxic ("HI")
• Oxygen ("HI")
• The Peak/Hi and Min/Low values can be cleared.
• When this page is displayed, press the TEST button to clear.
"CLR" displays.
4) Short Term Exposure Limit, STL.
The STL reading displays that the instrument calculated since
power-ON.
When this page displays, press the TEST button to clear.
"CLR" displays.
STEL is automatically reset to zero when the instrument is
turned ON.
STEL value is calculated over a 15-minute exposure.
The following formula is used to calculate the STEL value:
If the instrument is ON for less than 15 minutes:
the balance of the minute PPM values is set to zero and
the total is divided by 15 minutes.
Quick changes in barometric pressure result in an
offset in the sensor reading for a few minutes. If this
occurs, do not use the instrument for a period of five
minutes; then, perform a Fresh Air Set-up (FAS) to
re-zero the sensor. Perform the FAS in fresh air only.
(Minute #1 PPM value) + (Minute #2 PPM value) +....
(minute #15 PPM value) / (15 minutes) = PPM STEL value
2-13