OPERATION
3.5.
Displaying the Instrument Data - excluding the Steel Mill Version
A single, quick press on the button turn ON the backlight for 10 seconds. The
Information page can be accessed by pushing the button for about one second.
!
The instrument beeps twice.
There are following modes:
1)
Test gas mode
2)
Minimum Oxygen concentration ("
3)
Peak Reading
!
Toxic ("HI")
!
Oxygen ("HI")
!
The Peak/Hi and Min/Low values can be cleared.
When this page is displayed, press the button to clear.
!
!
4)
Short Term Exposure Limit ("
The STL reading displays that the instrument calculated since power-ON.
When this page is displayed, press the button to clear.
!
!
STL value is automatically reset to zero when the instrument is turned ON.
STEL value is calculated over a 15-minute exposure.
The following formula is used to calculate the STEL value:
((Minute #1 PPM value) + (Minute #2 PPM value) +...
+ (Minute #15 PPM value)) / (15 minutes) = PPM STEL value
If the instrument is ON for less than 15 minutes:
!
!
5)
Time Weighted Average ("
The TWA reading displays what the instrument has calculated since power-ON.
When this page is displayed, press the button to clear.
!
!
TWA value is automatically reset to zero when the instrument is turned ON.
TWA value is calculated over an eight-our exposure.
The following formula is used to calculate the TWA value:
Sum of 1-minute gas readings/480 minutes = PPM TWA value
If the instrument is ON for less than 8 ours (480 minutes), the balance of the
minute ppm values is set to zero.
24
CLR
"
" is displayed
Value is deleted
CLR
"
" is displayed
Value is deleted
the balance of the minute PPM values is set to zero and
the total is divided by 15 minutes