Climate Control Manual
Note: Images depict the US product. Temperatures will be displayed in celsius for all other regions.
Cleaning and maintenance
You should regularly carry out cleaning and maintenance as described in this section.
Daily
Disconnect the air tubing and hang it in a clean, dry place until next use.
Notes:
•
Do not hang the air tubing in direct sunlight as it may harden over time and eventually crack.
•
Do not wash the air tubing in a washing machine or dishwasher.
Weekly
1. Remove the air tubing from the H5i and the mask.
2. Wash the air tubing in warm water using mild detergent.
3. Rinse thoroughly, hang and allow to dry out of direct sunlight.
4. Before next use, reconnect the air tubing to the air outlet and mask.
Technical specifications
Maximum recommended pressure
Inner diameter
Length
ClimateLine temperature range
ClimateLine temperature cut-out
Material
Electromagnetic compatibility
If enabled by the clinician, Climate Control can also be set to
Manual. In this mode, air temperature and humidity can be
set independently, providing you greater flexibility but without
the rainout protection guaranteed in Auto mode. In this mode,
air temperature (61–86°F (16–30°C)) and humidity level (0–6)
can be set independently. If rainout occurs, either increase air
temperature or decrease the humidity setting.
ClimateLine
25 cm H
O
2
0.6" (15 mm)
6'6" (2.0 m)
60°F–86°F (16°C–30°C)
≤ 106°F (≤ 41°C)
Flexible plastic and electrical components.
Refer to the S9 Information Guide.
/
MAX
ClimateLine
MAX
ClimateLine
Oxy
30 cm H
O
2
0.75" (19 mm)
6'3" (1.9 m)
60°F–86°F (16°C–30°C)
≤ 106°F (≤ 41°C)
English
3