MESSAGES OF THE ALARM AND ERROR DISPLAYS
If an error message is shown the sterilization cycle has failed. It is necessary
to repeat all the preparing and sterilization phases
If an error message appeared on the display during the sterilization phase, the cycle failed
and the instruments in the chamber were not sterilized.
To reset the error message push the
MESSAGE
The sterilization cycle was interrupted by the operator pressing the
ERROR100
START/STOP
ERROR120
Low pressure in the chamber while sterilizing
ERROR125
High pressure in the chamber while sterilizing
The pressure did not increase properly during the preheating phase or the
machine could not discharge the pressure at the end of the sterilizing cycle.
ERROR140
If this is happening even if the water is in the chamber, dry and empty the
sterilization chamber, reset the alarm, let the water in and restart the cycle
up
ERROR199
The power was disconnected while the sterilization cycle
The cycle was started up without let water in the chamber.
Dry and empty the sterilization chamber, reset the alarm, let the water in and
ERROR400
restart the cycle up.
The cycle was started up but 5 min went from the water inlet in the chamber.
ERROR401
Dry and empty the sterilization chamber, reset the alarm, let the water in and
restart the cycle up
The cycle was started up but the machine was hot. After the door
the machine let the water out of the chamber.
ERROR450
Reset the alarm, wait until when the machine is cold, Dry and empty the
sterilization chamber, let the water in and restart the cycle up
During the preheating phase, in 40 minutes was not reached pressure value
established for sterilization.
ERROR700
If this is happening even if the water is in the chamber, dry and empty the
sterilization chamber, reset the alarm, let the water in and restart the cycle
up
ERROR760
Failure in the device for the pressure level
ERROR800
The pressure did not go down properly during the drying phase.