The installation is a fundamental operation for the following use and good operation of the
device.
Here is a list of the points to be followed in order to achieve a good installation.
1. The device must be installed inside a laboratory accessible only to the authorised
personnel.
2. The room in which Autoclave is installed must be lighted according to standard UNI
10380 and enough ventilated.
3. Position the device on a plane surface with minimum capacity of 60 kg.
4. The autoclave feet are adjusted by the producer, in order to obtain a correct level on
plane. Their modifications could compromise the autoclave good operation Photo A.
5. Leave at least 5 cm between the device rear part and the wall.
6. Position the autoclave at such a height as to make it possible for the operator to check
the whole sterilisation chamber and to carry out the normal cleaning operations.
7. Do not install the autoclave near washing basins, taps, because the device furniture is
not waterproof.
8. Do not place trays, papers, containers with liquids, etc. on the device, because
the device grids are used to ventilate and must not be obstructed.
9. Do not lean on the door
10. Do not install the device near heat sources (for instance: Autoclave, Bunsen burner,
kiln); keep a minimum distance of 40 cm.
11. In case the filling tank is directly discharged in waste waters, position the device at a
height which is higher than the discharge.
12. The device cannot be built in because the water filling in the tank is carried out from
above.
13. Install the device in such as way as not to excessively bend the supplying cable.
INSTALLATION
1A
when it is opened.