Order List - Dräger Panorama Nova S Instrucciones De Uso

Ocultar thumbs Ver también para Panorama Nova S:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 26
Check the speech diaphragm and O-ring, and replace if
necessary.
Press the inner mask back into place. In doing so, ensure that the
entire edge fits in the groove. The centre markings on the
connector and the inner mask must line up.
Replacing the exhalation valve disc
Replace the exhalation valve disc (see "Visually inspect the
exhalation valve" on page 15).
Fit the exhalation valve cover. Ensure that it engages.
Replacing the inhalation valve or valve disc
Pull the connector out of the mask body (see "Replacing the
connector" on page 16).
Remove the old valve disk and fasten the new disk.
Insert the connector in the mask body (see "Replacing the
connector" on page 16).
Storage
Loosen the harness as far as possible (handle tab).
Wipe uncoated visors with an anti-static cloth.
Do not wipe coated visors.
Pack the breathing connection in its bag or storage box.
Store it in a dry, dust-free place where it is not deformed.
Storage temperature -15 °C to +25 °C.
Keep out of direct sunlight and heat.
Observe ISO 2230 "Rubber Products - Guidelines for Storage", and
national guidelines for the storage, maintenance and cleaning of
rubber products.
Dräger rubber products are treated with an agent to protect against
premature aging which in some cases is visible as a grey-white
coating. This coating can be washed off with a brush and soapy water.

Order list

Designation and description
Panorama Nova S – EPDM – PC – P
Panorama Nova S – EPDM – PC/CC – P – with
screen
Panorama Nova S – SI – PC – P
Mask spectacles
"klar-pilot" gel
Sekusept cleaner
Incidur
Mabox I carry-box
Mabox II carry-box
Part number
R 51 853
R 55 210
R 54 743
R 51 548
R 52 560
79 04 071
79 04 072
R 53 680
R 54 610
17
Tabla de contenido
loading

Este manual también es adecuado para:

Panorama nova p

Tabla de contenido