– The display containing the current gas concentration changes with the
display » OK «.
Press the [OK] key to exit the calibration function or wait for approx. 5 seconds.
If a fault has occurred during the fresh air calibration:
– The fault message »
« appears and »
respective sensor instead of the measured value.
In this case, repeat the fresh air calibration. If necessary, have the sensor
replaced by qualified personnel.
Calibrating/adjusting the sensitivity for an individual
measuring channel
– The span calibration can be carried out selectively for individual sensors.
– During the span calibration, the sensitivity of the selected sensor is set to the
value of the test gas used.
– Use a standard test gas.
Allowed test gas concentration:
Ex:
40 to 100 %LEL
O
10 to 25 vol. %
2
CO: 20 to 999 ppm
H
S: 5 to 99 ppm
2
Test gas concentrations of other gases: see Instructions for Use for the
respective DrägerSensors.
Connect the test gas cylinder with the calibration cradle.
Vent the test gas into a fume cupboard or into the open air (with a hose
connected to the second connector of the calibration cradle).
Never inhale the test gas. Health hazard!
Observe the hazard warnings of the relevant Safety Data Sheets.
Switch on the instrument and insert it into the calibration cradle.
Press the [ + ] key and keep it pressed for 5 seconds to open the calibration
menu, enter the password (password on delivery = 001).
Use the [ + ] key to select the single gas calibration function. The symbol for
span calibration »
« flashes.
Press the [OK] key to start the channel selection.
– The flashing display shows the gas of the first measuring channel,
e.g. » ch4 - %LEL «.
« is displayed for the
CAUTION
Press the [OK] key to start the calibration function of this measuring channel,
or use the [ + ] key to select another measuring channel (O
CO - ppm, etc.).
– The test gas concentration is displayed
Press the [OK] key to confirm the test gas concentration or use the [ + ] key
to change the test gas concentration and complete the process by pressing
the [OK] key.
– The measurement value flashes.
Open the test gas cylinder valve to let gas flow over the sensor with a volume
flow of 0.5 l/min.
– The displayed, flashing measurement value changes to the value according
to the supplied test gas.
When the displayed measurement value is stable (after at least 120 seconds):
Press the [OK] key to perform the calibration.
– The display containing the current gas concentration changes with the
display » OK «.
Press the [OK] key or wait for approx. 5 seconds to end the calibration of this
measuring channel.
– The next measuring channel may appear for calibration.
– After the calibration of the last measuring channel, the instrument changes to
measuring mode.
Close the test gas cylinder valve and remove the instrument from the
calibration cradle.
If a fault occurred during the span calibration:
– The fault message »
« appears and »
respective sensor instead of the measured value.
In this case, repeat the calibration.
If necessary, replace the sensor.
Notice for the adjustment of the ex-channel to nonane as a measuring gas:
–
When calibrating the Ex channel, propane can be used as a substitute for
the test gas.
–
When using propane to adjust the ex-channel to nonane, the display must
be set to twice the used test gas concentration.
Notice for use in underground mining:
–
When calibrating the Ex channel to methane as the measured gas,
the indication on the instrument must be set to a value 5 % (relative) higher
than the test gas concentration used.