Riester ri-vital Instrucciones De Uso página 76

Ocultar thumbs Ver también para ri-vital:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Body temperature measurement with the optional ear thermometer:
The device can optionally be equipped with a holder for the
in place of the predictive thermometer.
This holder allows convenient storage of the thermometer and the protective sleeves
required for the measuring head.
For more information about using the device and important instructions regarding
measuring precision, please refer to the operating manual for the
12.3 Blood pressure (NIBP)
Disinfectable bladderless cuffs: all cuff sizes from the Riester portfolio
Measurement range:
Measurement range
for neonates:
Measurement precision:
Leakage:
Overpressure limit:
Time per measurement
ri-vital
TECHNICAL DATA
76
Systole 25 - 280 mmHg
Diastole 10 - 200 mmHg
MAP 15 - 260 mmHg
Systole 20 - 155 mmHg
Diastole 5 - 110 mmHg
MAP 10 - 130 mmHg
Mean value difference < +/-5mmHg
Standard deviation < 8mmHg
< 6 mmHg /minute.
300 mmHg mode „Adult"
150 mmHg mode „Neo" (neonates)
Overpressure limit (2nd Level): 330 mmHg mode „Adult"
165 mmHg mode „Neo"(neonates)
approx. 25s on average „Adult" mode max. 170 s
„Neo" mode max. 80s
®
ri-thermo
Nthermometer
®
ri-thermo
N
Tabla de contenido
loading

Tabla de contenido