• In order to avoid cross-contamination, only use BLUE thermometer probes to measure
ORAL or AXILLARY temperature.
•
Only use RED thermometer probes for measuring RECTAL temperature.
•
Do not open
any parts that can be maintained by customers. Opening the device can impair the
calibration and will invalidate the warranty.
•
Disposal of used probe covers must comply with current medical practices and the
local regulations concerning disposal in medical facilities.
•
Further information on disposal of the lithium batteries is available in the respective
community or from the environmental advisory board responsible.
•
Used electrical and electronic devices should not be disposed of with normal household
waste, but should be disposed of separately according to national or EU guidelines.
•
This device is only intended for use by trained medical persons.
CAUTION:
•
The device and the finger sensor should be cleaned with a soft, clean cloth. Please
use white spirit or isopropyl alcohol to remove heavy soiling. Other agents, especially
acetone-based substances, damage the housing or the display.
•
Do not press the key with hard or sharp objects.
•
Only accessories intended by Riester may be used to operate the device.
•
A function tester cannot be used to test the measurement accuracy of the pulse
oximeter or its sensors.
ri-vital
WARNINGS AND PRECAOUTIONS
ri-vital
and/or the predictive thermometer. The device does not contain
49