Replacing The Fuses - PeakTech 3355 Manual De Instrucciones

Ocultar thumbs Ver también para 3355:
Tabla de contenido
Idiomas disponibles
  • ES

Idiomas disponibles

6. Replacing the fuses

WARNING!
To avoid electric shock, disconnect all the test probes before
removing the fuse. Replace only with the same type of fuse. Not
note remove the top cover. Service should be performed only by
qualified personnel.
CAUTION!
For continued protection against fire or other hazard, replace only
with fuse of the specified voltage and current ratings.
Follow these steps to replace the fuse:
1. Press ON/OFF button to turn the meter off and disconnect the
test probes.
2. Remove the Battery cover (screws B) and the battery.
3. Remover the back cover by unscrewing the 6 screws (A)
4. Remove the blown fuse. Use only a fuse with the proper size
and value 500mA/1000V (6,3x32mm) or 10A/1000V (10x38mm)
5. Replace the back cover, the Battery and the battery cover.
-58-
Tabla de contenido
loading

Este manual también es adecuado para:

3360

Tabla de contenido