65
Hinweise zur
elektromagnetischen
Verträglichkeit (EMV)
Recommended separation distances between portable and mobile RF communications equipment and the
The ENDOFLATOR
®
40 model UI400 is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
The customer or user of the unit can help prevent electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the unit as recommended below, according to the maximum output power of the
communications equipment.
Rated maximum output
power of transmitter
[W]
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using
the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to
the transmitter manufacturer.
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and re ection from structures, objects
and people.
Electromagnetic
Compatibility (EMC)
information
Table 6
ENDOFLATOR
®
40 model UI400
Separation distance d [m]
according to frequency of transmitter
150 kHz to 80 MHz
80 MHz to 800 MHz
D
D
0.12
0.38
1.2
3.8
12
Indicaciones sobre
compatibilidad
electromagnética (CEM)
800 MHz to 2.5 GHz
F
0.12
0.23
0.38
0.73
1.2
2.3
3.8
7.3
12