62
Hinweise zur
elektromagnetischen
Verträglichkeit (EMV)
Cables which have been shown to comply with IEC 60601-1-2:
Cable type
Shielded
PA
No
SCB Cable
Yes
Power Cord
No
Guidance and manufacturer's declaration – electromagnetic emissions
The ENDOFLATOR
The customer or user of the ENDOFLATOR
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage uctuations/ icker
emissions
IEC 61000-3-3
Electromagnetic
Compatibility (EMC)
information
Length [m]
Ferrite
> 3
No
> 1
No
3
No
Table 1
®
40 model UI400 is intended for use in the electromagnetic environment speci ed below.
®
40 model UI400 should ensure that it is used in such an environment.
Compliance
The ENDOFLATOR
Group 1
Therefore, its RF emissions are very low and are not likely to cause interference in
nearby electronic equipment.
The ENDOFLATOR
Class B
domestic establishments and those directly connected to the public low-voltage
power supply network that supplies buildings used for domestic purposes.
Class A
Complies
Indicaciones sobre
compatibilidad
electromagnética (CEM)
Used for
Potential equalization
connecting SCB network, other SCB units
Power supply
Electromagnetic environment – guidance
®
40 model UI400 uses RF energy only for its internal function.
®
40 model UI400 is suitable for use in all establishments including