Before use, carefully read this manual with the safety instructions.
The product must only be used by physicians trained in its use. Training material is available upon request from the
manufacturer.
If disconnecting from the oxygen source make sure to always disconnect at the gas probe.
The connecting tube is for single use and must not be reprocessed. Reprocessing influences the function of the con-
necting tube. Any reuse involves the potential risk for cross contamination.
The connecting tube must not be used if the packaging is damaged or the shelf life has elapsed.
INSTRUcTIoN FoR USE
1. Connect the Manujet III to the oxygen source.
2. Pull the pressure regulator knob and turn it counter clockwise until the pointer on the gauge is at 0 bar / psi. Release
the pressure in the system by pressing the trigger.
3. Connect the Manujet III to the inserted catheter via the small bore tubing.
4. Adjust the desired ventilation pressure by turning the pressure regulator knob clockwise. Start with the lowest pressure
(0 bar/psi) and afterwards increase this pressure slowly under ventilation and clinical control of chest movements. The
adjusted ventilation pressure is displayed on the gauge. Push the pressure regulator knob for locking. To re-adjust the
pressure pull the pressure regulator knob and adjust to the desired pressure.
5. By pressing the trigger the patient can be ventilated continuously. Ventilation pressure and ventilation frequency must
be adjusted properly to achieve balanced ventilation.
The following minute volumes are delivered to the patient considering an in- and expiratory ratio of 1 to 1: (This infor-
mation is only valid for elective use, open airways and for the use with VBM Jet Ventilation Catheters):
catheter
14 G
13 G
DEVIcE DELIVERy
manujet III standard
Manujet III (standard) is supplied with a 4 m (13 ft.) high
pressure hose assembly and a 100 cm (40 inch) small bore
tubing (with Luer Lock fitting).
Gauge
Pressure
regulator
Ventilation frequency [1/min]
20 - 30
12 - 13
Connecting tube
Luer Lock fitting
Trigger (On/Off Valve)
Gas probe
High pressure hose
Ventilation pressure [bar/psi]
2.0 / 29
3.5 / 51
Version
manujet III with NIST-o
Manujet III with NIST-O
assembly) is supplied with a 100 cm (40 inch) small bore
tubing (with Luer Lock fitting).
Not for sale in USA and canada.
-9-
Volume [l/min]
16 - 21
24 - 30
2
fitting (without high pressure hose
2
NIST-O
fitting
2