In order to check the values, the Luer Lock fitting of the Manujet III has to be connected to a calibrated pressure gauge
(measurement accuracy of min. ±0.016 bar / ≈0.232 psi). Inflate the Manujet III to 1, 2, 3 and 3.5 bar (51 psi) and take
a note of the values. The values must be in the tolerance of ±0.05 bar (±0.72 psi). If this is not the case, the device must
be returned to VBM for service. If the values are within the tolerance, the device can still be used.
If required measuring equipment is not available, the metrological control can be carried out by VBM. In this case the
Manujet III has to be sent to VBM.
Note for conversion of bar to SI-Unit Pascal: 1 bar (14.5 psi) = 100 kPa
ShELF LIFE
manujet III
Shelf life for Manujet III is 8 years.
connecting tube
Expiry date: See connecting tube label.
coNSUmABLES
REF
Length
54-05-111
100 cm
(40 inch)
DISpoSAL
Used or damaged products must be disposed in accordance with the applicable national and international regulations.
SERVIcE
The medical device may only be serviced by persons who have the technical knowledge, qualification and the required
means to properly carry out the service.
After service of the medical device, the essential constructive and functional characteristics for safety and functionality
have to be checked, insofar as they can be influenced by the service procedure.
All warranty claims will be lost, if the user or an unauthorised service agency have attempted to effect repairs which are
not described in this instruction of use.
To ease repair of the medical device, return it together with a detailed description of the defect.
As a protective measure for the safety of VBM staff return any medical device completely cleaned and disinfected (see
"Wipe Disinfection Manujet III"). The VBM Service is entitled to refuse the repair of contaminated items for safety reasons.
REpAIR
The glass of the gauge (REF 52-03-903) can be easily replaced using a screw
driver (a). The chapter „Metrological Inspection" must be considered.
mAINTENANcE
The Manujet III is maintenance-free if operated and handled properly.