5
Performance Tests and Calibration
Table 5-4
Calibration Adjustments (continued)
Step
Test Function
3
Turn the rotary switch to
4
Turn the rotary switch to
80
Cal Range
Short
position
1 V
Short
10 MΩ
position
1000 kΩ
100 kΩ
10 kΩ
1000 Ω
Keysight U1241B and U1242B User's and Service Guide
Input Reference Value
Dual banana plug with
copper wires short
between two terminals
1 V
Dual banana plug with
copper wires short
between two terminals
Input terminals open
(remove all test leads
and shorting plugs from
input terminals)
10 MΩ
1000 kΩ
100 kΩ
10 kΩ
1000 Ω
Cal Item
U1241B
U1242B
SHrt
1.000 V
SHrt
oPEn
10.000 MΩ
1000.0 kΩ
100.00 kΩ
10.000 kΩ
1000 Ω
loading