Performance Tests and Calibration
Calibration Adjustments
Review the
NOTE
beginning the adjustment procedures.
1 Turn the rotary switch to Test Function position, as shown in
2 After unsecuring the multimeter, the multimeter will go into the adjustment
3 The primary display will show the reference value of the Cal Items. Configure
4 Use the
5 Apply the input signal as shown in the Valid Input Reference Value column of
Users are highly recommended to complete the adjustments in the same order
NOTE
as shown in
6 Enter the actual applied input, see
7 Press
8 Upon completion of each adjustment value, the secondary display will show
If the adjustment failed, check the input value, range, function, and entered
NOTE
adjustment value and repeat the adjustment steps.
78
Test Considerations
mode, see
Unsecuring the Multimeter for
each Cal Item shown in Valid Adjustment Input Reference Values in
and
the
Table
5-3. The bar graph will display the input reading. There is no bar
graph display for temperature adjustment.
Table
5-3.
to start the adjustments. The CAL flashes in the secondary
display indicating the calibration is in progress.
PAS. If the adjustment failed, the multimeter will sound a long beep and the
calibration error number appears in the secondary display. The primary display
remains at the current Cal Item.
and
Adjustment Considerations
Calibration.
arrow keys to select the Cal Range.
Entering Adjustment
Keysight U1241B and U1242B User's and Service Guide
before
Table
5-3.
Table
5-3.
Values.
loading