Valid Adjustment Input Values
Adjustment can be accomplished using the following input values below.
Table 5-3
Valid adjustment input values
Function
Range
1000 mV, 10 V, 100 V, 1000 V
V
1000 mV, 10 V, 100 V, 1000 V
1 V
Ω
1000 Ω, 10 kΩ, 100 kΩ, 1000 kΩ, 10 MΩ
1000 nF, 10 μF, 100 μF, 1000 μF, 10 mF
1000 μA, 10000 μA
100 mA, 1000 mA
10 A
T1
0 °C
DCmV
100 mV
(T1)
The minimum AC current output of Fluke 5520A calibrator is 29 μA. Ensure
CAUTION
that at least 50 μA is set at the calibrator source of AC μA.
Keysight U1241B and U1242B User's and Service Guide
Performance Tests and Calibration
Valid Input Reference Values
0.9 to 1.1 x Full Scale
0.9 to 1.1 x Full Scale
0.9 to 1.1 x Full Scale
0.9 to 1.1 x Full Scale
0.9 to 1.1 x Full Scale
0.9 to 1.1 x Full Scale
0.9 to 1.1 x Full Scale
0.9 to 1.1 x Full Scale
Ensure that 0 °C with ambient compensation is
provided
0.9 to 1.1 x Full Scale