Step-by-Step Instructions
6.4

Replacement of the Rechargeable Battery

Propex Pixi
rechargeable battery.
If a fully charged battery is not sufficient for normal device operation
during at least one working day, the battery should be replaced by a
new one at your earliest convenience.
Please note that a new battery should be charged for 24 hours prior to
the first use of the device.
Warning
Use GP100AAAHC or compatible NiMH rechargeable battery only.
Usage of non-rechargeable batteries may cause device damage.
The battery compartment is located at the rear side of Propex Pixi
Apex Locator.
1. Carefully lift and tilt back the silicon cover to expose the screw.
Release the screw.
16/1346
®
Apex Locator is powered by one 1.2V AAA NiMH
Fig. 5
®