Follow-Up; After Each Treatment Session; At The End Of The Work Day - Dentsply Sirona Midwest E Plus Manual Del Operador

Ocultar thumbs Ver también para Midwest E Plus:
Tabla de contenido
Idiomas disponibles

Idiomas disponibles

Sirona Dental Systems GmbH
Operating Instructions Midwest E Plus
66 04 206 D3686
D3686.201.01.01.02 08.2016
6

Follow-up

6.1

After each treatment session

NOTICE! Condition immediately, or at the latest, one hour after treatment.
Immediately after treatment
NOTICE! Only use Midwest® Plus Aerosol Spray.
Using the T1 spray
The bur instrument must not be moving.
Instruments _ bur instrument
Wear appropriate protective clothing.
1. Purge the water and air channels on the treatment center for 30
seconds.
2. Remove the bur instrument with tweezers.
3. Pre-disinfect directly at the treatment center [ → 14].
4. Detach the instrument from the motor.
5. Transport the instrument to the hygiene room in a suitable transport
container.
6. Conduct automatic conditioning [ → 14]. Manual conditioning [ → 15] is
possible in exceptional cases if the national/local regulations are
followed.
7. Apply spray to the instrument [ → 15].
8. Sterilize the instrument and accessories [ → 17].
6.2

At the end of the work day

➢ Apply spray to the instrument [ → 15].
Instrument
NOTICE! Do not leave any instruments on the motor overnight, in order
Remove the motor at night
to prevent oil from leaking into the electric motor. Never lubricate the
electric motor.
6 Follow-up
6.1 After each treatment session
13
Tabla de contenido
loading

Tabla de contenido