Step-by-Step Instructions
6.6
Apex Localization
6.6.1
Getting Started
Disconnect the charger from the device if connected.
1. Before connecting the measurement cable with attached lip clip and
file clip to the patient, plug measurement cable into the device and
turn on the device by pressing the
of the device. The first bar will start blinking.
2. Attach the lip clip to the patient.
3. Gently insert the file into the canal.
Note
To ensure optimal performance the file size should be adjusted to the
canal diameter.
4. Connect the file clip to the metal shaft of the file.
The first bar will stop blinking accompanied by a double beep signal –
see
Fig.
Note
A blinking bar indicates faulty connection. Disconnect the
measurement cable from the patient and check cable connections,
clean the file clip and the lip clip, moisten the canal if necessary and
start again.
No other adjustments are required before starting apex localization.