7 Dismount
Error
Cause
Red sig-
Vibrating
nal lamp
frequency
flashes
out of speci-
(switching
fication
output high-
Buildup on
impedance)
the vibrating
element
Vibrating
element da-
maged
7
Dismount
7.1
Dismounting steps
Warning:
Before dismounting, be aware of dan-
gerous process conditions such as e.g.
pressure in the vessel, high temperatu-
res, corrosive or toxic products etc.
Take note of chapters " Mounting" and " Con-
necting to voltage supply" and carry out the
listed steps in reverse order.
7.2
Disposal
Pass the instrument on to a specialised recy-
cling company and do not use the municipal
collecting points.
Remove any batteries in advance, if they can be
removed from the device, and dispose of them
separately.
If personal data is stored on the old device to be
disposed of, delete it before disposal.
If you have no way to dispose of the old instru-
ment properly, please contact us concerning
return and disposal.
16
Rectification
Check the vibrating ele-
ment on buildup and
remove it
Check the vibrating ele-
ment and the sensor if
there is buildup and re-
move it
Check if the vibrating
element is damage or ext-
remely corroded
LFV200 • IO-Link