Check Canister and Tubing Connection
Indications That a Leak Condition Has Been Corrected
Therapy unit will become quiet, running only intermittently.
•
Audible leak alert will stop; visual alert will turn off. There may be a brief delay between
•
when the leak is corrected and the alert is discontinued.
The PREVENA™ Dressing will be compressed.
•
PREVENA™ 125 THERAPY UNIT DISPOSAL
At the end of therapy, the patient should return the PREVENA™ 125 Therapy Unit to the
physician for disposal. Dispose of all waste according to local requirements. Improper disposal
may run the risk of regulatory non-compliance.
1. Ensure canister is securely locked in the therapy unit. When
canister is installed, a distinct click will be heard indicating it has
been properly installed. Side tabs on canister should be flush with
unit.
2. Check tubing connectors to ensure they are fully engaged and
locked.
Dressing compressed - system pressure acceptable.
3
Dressing not compressed - system pressure not acceptable. Return
to the Correcting a Leak Condition section to continue pressure
correction steps.
X
17