MEGABRAS TR8703 Manual De Uso página 31

Medidor de relación de transformación
Idiomas disponibles

Idiomas disponibles

4.2. Running a test
4.2.1. VT/PT test
In this section, an example of running a test using the following
configuration (see § 4.3 for how to configure the instrument) is shown:
 Set Date/Time: July 4, 2010. 6:00 PM
 Set values in Nameplate:
Primary Voltage:19920; Secondary Voltage: 7200
 Select Type: VT/PT
 Select Test Mode: CONT/Ratio
 Storage Mode: Manual
 Set Filter: Normal
 Set Language: English
- Typical Connection
H RED
H BLACK
X BLACK
Alternate Connection
H RED
X BLACK
H BLACK
RATIO 2.767 : 1
PRIMARY
H RED
X RED
H BLACK
RATIO 1.383 : 1
PRIMARY
H RED
X RED
H BLACK
32
SECONDARY
X RED
19920:7200CT
X BLACK
SECONDARY
X RED
19920:7200CT
X BLACK
loading