Temperature Dry Well Calibrators/Micro Calibration Baths
CTD9100/CTB9100
Calibrator type
(heating/cooling)
CTD9100-165
CTB9100-165
with silicone oil 20 CS
CTB9100-165
with distilled water
Calibrator type
(heating)
CTD9100-450, -650S
up to 300 °C
300 °C up to 100 °C
CTB9100-225
with silicone oil 20 CS
200 °C up to 50 °C
50 °C up to 30 °C
CTB9100-225
with distilled water
90 °C up to 50 °C
50 °C up to 30 °C
If you have carried out settings for
these three settings, the calibrator/
micro calibration bath uses the
new values only when modifying
the set temperature or switching
the calibrator off and on again.
A further procedure is to switch
off the automatic control prior to
modifying parameters (see section
6.3.1), and to switch it on again
afterwards (see section 6.3.2).
The heating and cooling gradients and the duration
time can be set in the parameter level 'rEG.
This is achieved by
Pressing the P key for approx. 5 sec., the main
menu opens.
OPEr appears on the upper PV display.
LED SET flashes on the lower SV display.
Fig. 36: Operator menu OPEr
WIKA Operating Instructions Temperature Dry Well Calibrators/Micro Calibration Baths
Press the P key again, the group level opens.
Setting for
OPEr appears on the upper PV display.
"SLoF"
'SP appears on the lower SV display and LED
SET flashes.
< 5 °C/min
< 6 °C/min
< 4 °C/min
Setting for
"SLoF"
Fig. 37: Group 'SP
Use the key to select the group 'rEG.
< 10 °C/min
OPEr appears on the upper PV display.
< 5 °C/min
'rEG appears on the lower SV display and LED
SET flashes.
< 4 °C/min
< 0.5 °C/min
< 2 °C/min
< 0.5 °C/min
Fig. 38: Group 'rEG
Press the P key again, the parameter level
opens.
'rEG appears on the upper PV display.
SLor flashes on the lower SV display.
Fig. 39: Parameters for heating gradient SLor
GB
21