13.4.2 CTB9100-225
Measuring conditions:
All heating and cooling periods are related to a
reference Pt100 probe, Ø 6 mm, stainless steel
sheath
The reference sensor is located 5 mm above the
protective basket in the middle of the tank
Distilled water
Heating up:
25 °C to 40 °C
40 °C to 50 °C
50 °C to 75 °C
75 °C to 90 °C
Silicone oil 5 CS
Heating up:
25 °C to 40 °C
40 °C to 50 °C
50 °C to 75 °C
75 °C to 100 °C
100 °C to 120 °C
Silicone oil 10 CS
Heating up:
25 °C to 40 °C
40 °C to 50 °C
50 °C to 75 °C
75 °C to 100 °C
100 °C to 125 °C
125 °C to 150 °C
150 °C to 165 °C
WIKA Operating Instructions Temperature Dry Well Calibrators/Micro Calibration Baths
Period
0:55 min
0:37 min
1:27 min
1:30 min
Period
0:51 min
0:16 min
0:54 min
1:13 min
1:35 min
Period
0:52 min
0:22 min
0:52 min
0:53 min
0:59 min
1:12 min
1:03 min
All heating and cooling periods do not include
stabilisation time
The measurements were carried out at a room
temperature of approx. 23 °C with bath without
cover
Cooling down:
90 °C to 75 °C
75 °C to 50 °C
50 °C to 40 °C
40 °C to 25 °C
Cooling down:
120 °C to 100 °C
100 °C to 75 °C
75 °C to 50 °C
50 °C to 40 °C
40 °C to 25 °C
Cooling down:
165 °C to 150 °C
150 °C to 125 °C
125 °C to 100 °C
100 °C to 75 °C
75 °C to 50 °C
50 °C to 40 °C
40 °C to 25 °C
Period
5:53 min
15:17 min
10:50 min
45:26 min
Period
3:27 min
5:55 min
10:00 min
7:02 min
34:28 min
Period
1:40 min
3:17 min
4:14 min
5:59 min
9:59 min
7:00 min
31:40 min
35