Temperature Dry Well Calibrators/Micro Calibration Baths
CTD9100/CTB9100
Fig. 4: Component description
WIKA Operating Instructions Temperature Dry Well Calibrators/Micro Calibration Baths
Front and top
At the top you will find the dry well access opening
for the inserts for models:
∅ 28 mm x 150 mm
CTD9100-165:
∅ 60 mm x 150 mm
CTD9100-450:
∅ 28 mm x 150 mm
CTD9100-650:
The controller with display and the four button
control pad are positioned at the front.
Rear
At the rear you will find the type label with important
information about the individual model:
CTD9100-165: range
CTD9100-450: range
CTD9100-650: range
The suitable line voltage and frequency is printed:
115 VAC, 50...60 Hz or 230 VAC, 50...60 Hz
100 VAC...240 VAC, 50...60 Hz
Further you will find the individual serial number, e.g.
S/N 550 33 44, the instrument's power consumption
and the fuse rating.
Bottom
Mains connector, power switch and fuse holder are
positioned at the bottom of the instrument close to
the front.
Furthermore the fans air inlet(s) can be found
there - one or two - depending on the model.
Do not obstruct air inlets.
-30 °C ...+165 °C
40 °C ... 450 °C
40 °C ... 650 °C
5
GB